Functional description
CANBus
A4461-2.0
HBM: public
49
Type = MeasVal
DataFormat = REAL32
Factor = 1.0
Offset = 0.0
StartBit = 0 bzw.
7 (depending on
„BitSequence“)
SignalLength = 32
BitSequence = INTEL or MOTOROLA
CanMessage TransmissionType = “Control” or
“Timer”
A received CAN message cannot be forwarded via CAN.
The signal source must also be taken from within the
module, that is:
ModuleReference (empty or “UUID=[own UUID]”),
SignalReference = “AnalogIn_Connector[2…8].Sig
nal1”
The parameters from <ValueOnError> are supported.
For timer-controlled transmission, a measured value can
only be sent every 10 ms, or less often. Please note that
up to and including firmware version 4.4, the parameter
<Period> has been given in milliseconds. To ensure that
the future use of this parameter is the same as in the
MX471B, <Period> must be given in microseconds in the
MX840B as well in future, with a resolution of 1000 μs.
6.6
Data bit numbering systems
according to the vector DBC format
6.6.1
Numbering systems used in QuantumX /
SomatXR parameterization
The data for the measured values can be interpreted in
different numbering systems.
Summary of Contents for SOMATXR
Page 1: ...Operating Manual Bedienungsanleitung English Deutsch CANBus...
Page 3: ...Operating Manual Bedienungsanleitung English Deutsch CANBus Receive Transmit...
Page 65: ...QuantumX SomatXR and CCP XCP on CAN CANBus A4461 2 0 HBM public 63...
Page 74: ...QuantumX SomatXR and CCP XCP on CAN 72 A4461 2 0 HBM public CANBus...
Page 75: ...Operating Manual Bedienungsanleitung English Deutsch CANBus Empfangen Senden...
Page 141: ...QuantumX SomatXR und CCP XCP on CAN CANBus A4461 1 0 HBM public 67...
Page 149: ...QuantumX SomatXR und CCP XCP on CAN CANBus A4461 1 0 HBM public 75...
Page 151: ...QuantumX SomatXR und CCP XCP on CAN CANBus A4461 1 0 HBM public 77...