8
TEDS
A1813-1.1 en
HBM
Some of the data in the templates can be stored by the user themselves, e.g.
- a measuring point identification number,
- filter settings,
- a zero value, a tare value,
- a short comment.
Detailed descriptions for individual templates can be found in Chapter 4.
User rights hierarchy
Various user rights are needed in order to amend the different entries in the
templates, and these rights may differ from one entry to the next within a
template.
•
Standard rights (USR level)
This level concerns rights that the operator of the transducer needs in order
to change entries which depend on the conditions of use.
•
Calibration rights (CAL level)
This level concerns rights which are needed by a calibration laboratory, for
instance, if the sensitivity in the TEDS memory needs to be changed.
•
Administrator rights (ID level)
Administrator rights in relation to TEDS are intended for the sensor
manufacturer. In the case of a self-built sensor or subsequent upgrading
with TEDS, HBM customers can of course also make use of these rights.