Changing the default SYNC-Frequency of 50 Hz, has to be made before ad-
justment!
•
Do a zero calibration
<CAL> <ZERO>
See page 62 for the flow chart. The process time is depending on the
selected average filter setting. It takes approx. 15 sec at filter 10.
•
Do a span calibration
<CAL> <SPAN>
See page 62 for the flow chart. The process time is depending on the
selected average filter setting. It takes approx. 15 sec at filter 10.
•
Select further settings in the OPTION menu
•
Enter/select further settings
4.3 Starting the digital Setup
Digital setup and calibration is carried out entirely from the front panel using
the secondary functions on the Control Buttons. There are two methods of
entering the digital setup program. The first setup access method is by
removing the security screw at the right of the keypad, and pressing the hidden
button behind the screw. This setup program allows all steps to be accessed,
including calibration.The second method involves pressing the Gross/Net
button for at least 2 seconds. This also allows access to the setup, but all
critical steps controlling the calibration and trade operation of the WE2110
cannot be accessed. This allows all other digital functions to be accessed
without risk to the basic calibration. Once configured and calibrated, the
WE2110 can be pass-code protected to prevent un-authorized tampering. If the
scale has been previously pass-code protected, the correct code will need to
be entered before setup can start.