66
WE2110
HBM
A2181-1.0 de/en
12
Error messages
Weighing errors
(U
−
−
−
−
)
The weight is below the minimum allowable weight reading
(O
−
−
−
−
)
The weight is above full range
(ZERO)
(ERROR)
The weight reading is beyond the limit set for Zero operation
(STABLE)
(ERROR)
Scale motion has prevented a Zero, Tare or Print operation from occurring on
command.
(PRINT)
(ERROR)
A printer problem has prevented the printout from being completed. (No power,
no paper, cable fault)
(CAL) (DUE)
Maintenance is due ! Call your scale supplier
(PRINT)
WE2110 is currently printing
(SAVING)
WE2110 is currently writing to the EEPROM
Setup errors
(RES) (LO)
The scale build is configured for less than 100 verification intervals
(RES) (HIGH)
The scale build is configured for more than 100,000 verification intervals
(CHEC)
(TRADE)
! Note:
This error is
not displayed
if parameters
are set via the
interface !
At least one parameter is not acc. to EN45501 requirements. Check the trade
relevant items. This is only displayed in trade mode when leaving setup.
Error numbers:
1 = Verification interval e1 or e2 > 50
2 = Graduations > 6000d
3 = No weighing unit selected
4 = Motion set to NONE
5 = Zero tracking not OFF or 0.5d/s
6 = Zero setting range not 2% or
−
1% +3%
7 = Zero dead band not set to 000000
8 = Buttons set for immediate operation
9 = Type set to direct mV/V calibration
Calibration errors
(ZERO) (HIGH)
The load cell output is beyond allowable zero calibration range.
(ZERO) (LO)
The load cell output is below allowable zero calibration range.
(SPAN) (LO)
The load cell signal range (span) is too small for these settings.
(SPAN) (HI)
The load cell signal range (span) is too large for these settings.
(NO) (ZERO)
A new WE2110 must be zero calibrated before span calibration.