Supervisormode
138
A4774-1.0
HBM: public
WTX110
10.2.10 Alarm (
Alarm:Empty
)
As soon as one of the alarms is set to
On
the relevant alarm is active. If none
of the alarms is indicating an error, the alarm status is cleared (“0”). If at least
one of the alarms is indicating an error, the alarm status is set (“1”). If a digital
output (Alarm) has been configured, it is set or reset according to the alarm
status.
Menu
Key
Function/Secondary selection
Alarm:Empty
Empty Off/On
Monitoring of the empty
weight at the start of dosing
(
Flow monitor Off/On
Activates the alarm for fill flow
monitoring (
see also sec
and
)
Overflow Off/On
Alarm in case of over/under
flow of gross signal, net signal
or ADC value
Dos.time Off/On
Alarm when the maximum
dosing time is exceeded (
see
Container Off/On
Alarm in case of deduction
weighing and gross measured
value < empty weight or gross
measured value < target
weight
Min.start Off/On