Calibration
170
A4774-1.0
HBM: public
WTX110
Low Capacity
999999
Enter the maximum load of the lowest range.
Example:
Low Capacity 1500
If the
Low Capacity
value is greater than
Mid
Capacity
, the error message
Entry Not Valid!
is
displayed.
Low Int.
999.9999
Enter the scale interval of the lowest range.
Example:
Low Int. 1.0000
The error message
Entry Not Valid!
is
displayed if you enter an invalid scale interval.
Menu
Key
Function/Secondary selection
Unit
kg
Specify the unit of measurement symbol
kg
kilogram
g
gram
t
metric ton
lb
pound
Back to
Select Group
11.5
Calibration
For more information on the keys necessary to access this menu group, go to
the
.
The scale is calibrated with this menu group. In addition to calibration at full
load, part load calibration is also possible. If you want to move the scale to a
different location after calibrating it, you can use the geo value to compensate
for the differing gravitational pull, without having to recalibrate the scale (not for
applications subject to weights and measures legislation). Precalibration
without weights is also possible by entering the standardized signal (mV/V).
Example of calibration of a tank weigher:
Summary of Contents for WTX110
Page 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Page 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Page 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Page 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Page 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...