Calibration
174
A4774-1.0
HBM: public
WTX110
Information
The Clr key can be used to clear the value and enter a new one.
→ If the internal resolution is too low, the following warning is displayed
Resolution Error
The internal resolution should be at least 10
times the set resolution.
Check calibration data and repeat calibration if
necessary.
Menu
Key
Function/Secondary selection
Select Group
Save Parameters?
Y
Parameters are saved
N
Parameters are not saved
The calibration values are saved when you exit
the
Select Group
, step, provided you answer the
Save Parameters?
prompt with
Y
Information
To discard the calibration values, you must exit Setup without saving (
Save
Parameters No
) before you can call another group.
11.6
Linearization
For more information on the keys necessary to access this menu group, go to
the
.
Linearization compensates for any variation of the scale characteristic (from
the ideal straight line) by setting up to 6 correction points adapted to the actual
scale characteristic.