Calibration
WTX110
A4774-1.0
HBM: public
177
Menu
Key
Function/Secondary selection
Select group 1-9
Press 4 x for group
4 Zero Adjust
→ Linearization points present
Fixpoints active!
Linearization entries have already been made;
they may influence the calibration.
Menu
Key
Function/Secondary selection
Unload Scale
Unload the scale.
Remove any auxiliary frame or other initial load.
When you press the Enter key the zero point is
measured.
Zero adjustment
Adjusting...
Measurement of the zero point
Zero: 0.00
Display the new zero point with 10 times higher
resolution (as a check).
Zero(mV/V):
0.50000
Display the standardized signal - see
Calculate
Span
Back to
Select Group