Calibration
WTX110
A4774-1.0
HBM: public
179
Information
The filtration level can also be changed retrospectively in
ServiceMode
and
Master mode
. This makes it possible to change the filtration level even after
legal verification, without breaking the verification seal.
The filter setting in
Service mode
overwrites the calibration value every time
the application is started. So to check the validity of the setting the value in
ServiceMode
must be checked!
Menu
Key
Function/Secondary selection
Auto Zero Range
0.5D
Automatic zeroing
This sets the range around the zero point in
which the weight value must be in order for
automatic zeroing to activate.
OFF
Turn off zeroing
0.5D, 1.0D, 3.0D
Range size
PbZero (%) +
999
Range of zero setting (+)
This sets the range above the power-up zero
point in which the scale can be zeroed by the
Zero key, and in which automatic zeroing takes
effect.
PbZero (%) -
999
Range of zero setting (-)
This sets the range below the power-up zero
point in which the scale can be zeroed by the
Zero key, and in which automatic zeroing takes
effect.
Power-up Zero +/- 10 %
Startup range of zero setting
This sets the range around the calibrated zero
point in which the scale is automatically zeroed
on power-up.
OFF
Power-up zero off.
±2 %, ± 10%
Range size
Summary of Contents for WTX110
Page 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Page 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Page 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Page 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Page 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...