Calibration
WTX110
A4774-1.0
HBM: public
183
Function/Secondary selection
Key
Menu
Maximum Y 15.000
Enter the maximum angle in the Y direction for
which the correction will still be performed. If the
value is exceeded the weight indicator is hidden.
Incline mode Correct
Operating principle of the optical tilt sensor:
Correction
The weight is corrected according
to the inclination.
Check
:
A check determines whether the
angle is within the valid range. The
display is hidden outside of this
range. There is no correction
(function like the tilt switch).
11.9.2
Incline calibration
Menu
Key
Function/Secondary selection
Incl.
calibration
0-Adj.Incl.Sensor?
Linearization of the optical tilt sensor with
unloaded scale
N
No
Y
Yes
X= 0.754 Y= -0.110?
Display of the current X, Y angle
Measure ...
Measurement running. The mean values of
multiple measurements are determined.
Offset X0
0.706
Display/enter the correction value in X direction