Calibration
184
A4774-1.0
HBM: public
WTX110
Function/Secondary selection
Key
Menu
Offset Y0
-0.126
Display/enter the correction value in Y direction
Determine Preload?
N
The weight of the preload is also required for the
correction. It must also be corrected. If the
preload is not known it can be determined here.
Measurements in straight and inclined position
are required for this.
Y
Yes
N
No
→ When determining preload
Determine Preload = Y
X=
0.754 Y= -0.110?
Display of the current X, Y angle
Start of the first measurement on an even level
Measure ...
Measurement is performed.
Change inclination
The second measurement should be performed
on an inclined level if possible
X=
0.754 Y= -0.110?
Display of the current X, Y angle
Measure ...
Measurement is performed.
Preload 200.0
Display/input of initial load
11.9.3
Incline linearization
Menu
Key
Function/Secondary selection
Incl.
Linearization
Linearization of the optical tilt sensor with partial/
full load
Linear
+X Y=0
Select the linearization direction