Calibration
WTX110
A4774-1.0
HBM: public
185
Function/Secondary selection
Key
Menu
+X Y=0
Linearization of the +X side
—X Y=0
Linearization of the -X side
X=0 +Y
Linearization of the +Y side
X=0 —Y
Linearization of the -Y side
Lin.Load(X=0 Y=0)?
N
Determine the weight without the effect of the
optical tilt sensor. The scale must be placed on a
straight level to do this.
Y
Yes
N
No
Load(X=0 Y=0) = Y
X= 0.754 Y= -0.110?
Display of the current X, Y angle
Mount load and confirm; continue with
measurement on straight level
Linear.(X=0 Y=0) ...
Measurement is performed.
Lin.Load
100.0
Display/input of linearization weight
X= 0.754 Y= -0.110?
Display of the current X, Y angle Position the
scale at the inclination to be calibrated (here +X).
Linear.(+X Y=0) ...
Measurement is performed.
Weight:
100.01
Display the weight value in 10 times higher
resolution (as a check of the scale)