PanelX interface
192
A4774-1.0
HBM: public
WTX110
12
PanelX interface
An Ethernet-based data connection to the WTX110 can be set up via the
PanelX interface. This data connection via port 443 is protected with standard
HTTPS encryption.
The connection is made through the LAN connection on the bottom of the
device.
The HBM PanelX communicates via this open interface with the WTX110,
thereby making it possible to calibrate and configure the WTX110.
Important
After the IP address of the WTX110 is changed, the SSL certificate of the
PanelX interface must be updated. Go to the Network menu (see section 6.7)
in Service Mode and select the Update SSL cert function. After the certificate
update is complete restart the device.
Make certain before you do this that the date and time of the WTX110 are set
correctly (see Supervisor mode/General, chapter 10, page 132).
Information
If there are problems setting up the connection, note the settings in the
Network menu (see section 6.7 "Netzwerk (Network)", page 89).