Continuous output/HBM protocol (
Cont.out
)
196
A4774-1.0
HBM: public
WTX110
Example of remote indication data set control
1st character: Start
character
Always S
'S 10.98 t
C
R
L
F
'
2nd character: Status
(Blank) = scale stable
'SD 10980 kg
C
R
L
F
'
D = scale in motion
3rd - 12th character
Weight 10 characters,
in format of scale calibration
13th character
Always “_” (Blank)
14th - 15th character
Unit of weight
16th - 17th character
CR and LF
Tab. 14.1 Example for customized protocol "083m :DN10 U013010"
14.1
HBM Remote protocol
This data set has a fixed format The following table describes the content of
the data set.
' 100.0 kgN
C
R
L
F
'
1st character: Status
~ = scale in motion
° = scale stable
1,2,3 = range of scale
2nd - 9th character
Weight 8 characters, in format of
scale calibration
10th character
Always
(Blank)
11th - 12th character
Symbol of units of measurement
13th symbol
N = net weight
(Blank) = gross weight
14th - 15th character
CR and LF