Transport, maintenance and cleaning
WTX110
A4774-1.0
HBM: public
203
Notice
Danger of short circuit The battery's positive terminal must not come into con
tact with the housing when replacing the battery!
Notice
Take proper measures to prevent electrostatic discharge. If you are not certain
what to do please contact HBM Service.
Important
Replace the batteries of legally verified calibrated systems every time a new
calibration is carried out!
The lithium battery backing up the RAM and the real-time clock will last at least
3 years in normal operation. The battery must be replaced by a trained service
technician every 3 years.
Replacing the type CR2032 battery
Information
Insert a new battery within 30 seconds, otherwise the stored data will be lost!
1. Cut the power to the device and make certain the power supply is no longer
connected!
2. Unscrew the cover plate on the back.