16.2
Error messages
In the event of an error during calibration or in weighing operation, the following
messages may be displayed:
Error message display
Possible causes
Remedy
ADM faulty
A/D converter not
supplying data
Replace A/D converter
Short-circuit in load cell
cable
Check wiring
EMC interference
Check wiring
ADM not factory cal.
Invalid Setup data
Calibration faulty
Repeat calibration
ADM faulty
Replace ADM
ADM not installed
ADM not installed
ADM not correctly
inserted
Check whether the A/D
converter is installed
Insert ADM (correctly)
Calibration locked
Error Calibr. Jumper
Jumper for calibration
lock in locked position
Remove bridge
I/O error
Interface error
Check interface and
replace if necessary
Incline sensor
Optical tilt sensor reporting error
Inclination too great
Change inclination
Optical tilt sensor faulty
Replace optical tilt sensor
Not available
No scale selected
Select scale in Service
mode