Interference
210
A4774-1.0
HBM: public
WTX110
Remedy
Possible causes
Error message display
Out of Range
A/D converter over/undermodulating because:
Load cell incorrectly
connected
Check wiring
and measurement signal:
0–20 mV/AC (signal)
approx. 5 V/AC
(excitation)
Load cell faulty
Check load cell
Extreme overload on
scale
Unload scale
O v e r l o a d
Scale overloaded
Unload scale
--------
Zeroing or taring cannot
be performed because
the scale is in motion
Stabilize the scale
CPU not receiving data
from weighing interface
Check external and
internal wiring
U n d e r l o a d
Gross weight values less
than
*
20 d (below zero)
Load the scale
Set parameter 'Underload
20d' to N=Off
Power-up Motion
Power-up motion.
This message appears
immediately after power-
up if the scale does not
find a stable weight value
within the set power-up
zero range (±2 %,
±10 %).
Stabilize the scale