Installation
WTX110
A4774-1.0
HBM: public
27
► Guide the cable with plastic insert into the metal housing.
► Screw on the swivel nut and tighten it with a screwdriver.
WARNING
Cut the individual wires to length so that the parts conducting mains voltage
(connection cable, power supply unit) cannot touch! Use wire end ferrules for
leads with fine wires and make certain no individual wires are sticking out.
4.5
Connection of analog SG sensors
The WTX110-A permits the connection of analog SG sensors with the following
specification:
S
Maximum 8 SG load cells with 350
Ω
each
S
Load cell impedance range 43
Ω
... 3321
Ω
S
Legal for trade weight resolution 10.000 e, internal 524.000 d
S
Smallest permissible input signal for legal for trade applications: 0.33 μV/e
S
Sample rate 50–800 measurements / second (can be adjusted in Service
mode)
S
Supply voltage for load cells: 5 V ±5 %, clocked
S
Connection in 4- or 6-wire configuration