Service mode
82
A4774-1.0
HBM: public
WTX110
Menu
Key
Function/Secondary selection
Config.
Analog out
AOut 1:
DAU15
DAU15
Internal analog output 15-bit
None
Not assigned
Notice:
If the device variant has an analog output,
DAU15 must always be selected!
Aout1: Mode
Select operating mode:
0-10 V, 2-10 V, 0-20 mA, 4-20 mA
→ Gross or Net selected
AOut 1: Scale
1
Selection of scale for output of gross/net weight
AOut 1:Calibration
N
Fine adjustment of the minimum and maximum
output signal:
N
No fine adjustment
Y
Fine adjustment using a multimeter
→ Perform calibration
AOut 1: 0V =
9
Calibrate zero point, for example =0V, change
analog signal incrementally
AOut 1: 10V =
4095
Calibrate full load value, for example =10V,
change analog signal incrementally