Conditions on site
Z16A
A04000_03_Y00_00
HBM: public
9
5
Conditions on site
Series Z16A... load cells are hermetically encapsulated and are therefore not
at all sensitive to the influence of moisture and humidity. The transducers
achieve protection class IP68 (test conditions: 100 hours under 1m water
column) and and IP69K (water at high pressure, steam cleaner), as per DIN
EN60529. Nevertheless, the load cells must be protected against constant
humidity and moisture.
Protection against corrosion
The load cell must be protected against chemicals that could attack the
transducer body steel.
Notice
Acids and all substances that release ions also attack stainless steels and their
welded seams.
Should there be any corrosion, this could cause the transducer to fail. If this is
the case, you must provide appropriate means of protection.
Cable protection
In applications where the TPE cable of the load cell can be chemically or
mechanically attacked or destroyed, protect the cable additionally by means of
standard cable protection systems. The load cell is fitted with an adapter that
has an outer diameter of 10mm to attach flexible cable protection tubing.
Mount the protection tubing with a cable tie or a hose clamp on the adapter.
Explosion protection version option
The ambient temperature range specified on the transducer must not be
exceeded.
Summary of Contents for Z16A
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z16A...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z16A...
Page 25: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z16A...
Page 47: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z16A...
Page 70: ...Dimensions 24 A04000_03_Y00_00 HBM public Z16A...
Page 71: ...Dimensions Z16A A04000_03_Y00_00 HBM public 25...