Electrical connection
16
A04000_03_Y00_00
HBM: public
Z16A
Electrical and magnetic fields often induce interference voltages in the
measuring circuit. Therefore:
S
Use shielded, low-capacitance measurement cables only (HBM cables fulfill
both conditions).
S
Connect the shielding extensively, not punctiform (connect over large
surface area).
S
Do not route the measurement cables parallel to power lines and control
circuits. If this is not possible, protect the measurement cable with e.g.
steel conduit.
S
Avoid stray fields from transformers, motors and contact switches.
S
Do not ground the transducer, amplifier and indicator more than once.
S
Connect all the devices in the measuring chain to the same protective earth
conductor.