Conditions on site
4.1 Ambient temperature
The effects of temperature on the zero signal and on the sensitivity are
compensated. To achieve optimal measurement results the nominal
temperature range must be maintained. Temperature-induced measurement
errors can be caused by uneven cooling or heating (for example by radiant
heat). A radiation shield and all-round heat insulation bring about marked
improvements. They must not form a force shunt.
4.2 Moisture
Extreme humidity or a tropical climate should be avoided if this means that the
classified limit values are exceeded (degree of protection IP50 under DIN EN
60529).
4.3 Air pressure
Changes in air pressure have no effect on the force transducer since it is not
hermetically sealed.
4.4 Chemical effects
The aluminium enclosure of the transducer is protected by a powder coating.
If used in difficult environmental conditions (direct weather effects, contact
with media which encourage corrosion) additional protective measures should
be employed by the user.