39
Z4A
HBM
A0697-7.0 de/en/fr
ZKM force introduction part
for precision measurement, for example DIN
EN10002-3-calibrations.
For use as a transfer standard or control standard for the Z4A force
transducer, we recommend using ZKM force introduction parts according to
DIN EN 10002-3 or ISO/FDIS376.
∅
D
C
B
ZKM tensile force introduction accessories
Dimensions apply to tensile force introduction
Type
ZKM
order number
Weight
(kg)
B
C
∅
D
min.
max.
Z4A/20kN
Z4A/20kN/ZKM
0.82
325
228
276
35
−
0.120
−
0.280
Z4A/50kN
Z4A/50kN/ZKM
1.45
350
248
299
45
−
0.130
−
0.290
Z4A/100kN
Z4A/100kN/ZKM
2.32
395
277
334
50
−
0.130
−
0.290
Z4A/200kN
Z4A/200kN/ZKM
4.19
447
317
382
64
−
0.170
−
0.330
Z4A/500kN
Z4A/500kN/ZKM
20.1
623
432
522
90
−
0.170
−
0.390
Installing:
•
Screw the thread of the ZKM force introduction part into the transducer’s
inside thread until it reaches the stop, then loosen it by one turn.