Mechanical installation
Z6R…
A4514-1.0
HBM: public
11
4.2
Mounting and load application
Attach the load cells at the mounting holes and apply
load to the other end. The screws and tightening torques
to be used are given in the following table:
Maximum
capacities
Thread
Min. property class
20 - 200 kg
M8
10.9
34 N
⋅
m
1)
Recommended value for the specified property class. Please comply with the screw
manufacturer's instructions with regard to screw dimensions.
Important
Load must not be applied to the side where the cable
connection is located, as this would cause a force shunt.
The load must act as accurately as possible in the direc
tion of measurement. Torsional moments, eccentric loads
and lateral or side forces cause measurement errors and
can permanently damage the load cell. Catch such inter
ference effects, e.g. using lateral stay bars or guide
rollers, but ensure that these elements do not absorb any
load or force components in the direction of measure
ment (force shunt which will in turn lead to measurement
errors).
To minimize error effects from load application, HBM
offers different load application elements, according to
the mounting conditions:
S
Pendulum bearing ZPL (Z6/200kg/ZPL) (for nominal
rated loads 20 kg … 200 kg)
S
Knuckle eyes ZGWR (Z6/200kg/ZGWR) (for nominal
rated loads 20 kg … 200 kg)
Summary of Contents for Z6R Series
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z6R...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z6R...
Page 31: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z6R...
Page 59: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Fran ais Z6R...