connect the measurement cable for the electronics
extensively to the junction box and extensively to the
shielding electronics housing.
S
The shield of the connection cable must not be used
for discharging potential differences within the system.
You must therefore lay sufficiently dimensioned
potential equalization lines to compensate for possible
potential differences.
Important
Potential equalization is specified for applications in
potentially explosive atmospheres.