To start the unit.
1.Place the unit on a solid, flat, dry surface making sure that the water tank and medication cup have been
properly filled.
2.Press the ON/OFF button.
To adjust the airflow, and nebulizer rate
Adjust the air flow volume by rotating the filter holder at rear.
Fig.5
To begin treatment
1.Follow the instructions provided by your physician and/or respiratory therapist.
2.Your Nebulizer is designed to start at a lower atomization rate in the beginning and will become stronger
after a few minutes.
Remember:
Keep your eyes away from the output of medication mist
If any discomfort or abnormality occurs, stop using the unit immediately.
Do not tilt or shake the unit when in operation.
Opening the top cover during operation may cause a severe injury.
To end a treatment
1.When the treatment has ended, press the ON/OFF button to shut off the unit. Otherwise, your unit will
automatically shut off after approximately 10 minutes of continuous use.
2.Unplug the adapter from the outlet and detach the adapter plug from the unit’s AC port.
3.Allow the unit to cool
– approximately 20-30 minutes.
Remember:
The unit should not be used for more than 30 minutes at a time.
Wait 30 minutes before reusing the unit.
After every use
1.Allow the unit to completely cool.
2.Remove the top cover .
3.Carefully remove the medication cup and pour out remaining medication
4.Empty the water tank.
5.Wipe out the water tank and transducer with a soft, lint-free cloth.
6.Follow the cleaning procedures provided in this guidebook.
Troubleshooting
If any abnormality occurs during use, please check and correct the following:
CONDITION
Unit does not operate when ON/OFF button is
pressed
No misting or low rate of misting
CORECTION
Check the AC adapter connection to the outlet
and/or check the AC plug to the unit’s AC port
Check whether the water tank is filled to the correct
level.
If using with optional battery pack, check whether the
battery is completely charged
Check whether there is medications in the
medication cup.