background image

 

16

 

Disconnect the charger power cord from the wall outlet when the batteries are 

fully charged. (See Fig 20)     

Important! 

Do not use for voltage input except the specified. Make sure your present voltage 

input (110V or 220V) and adjust manually. (See Fig 21)   

 

 

 

 plug-in controller  plug-in wall outlet          (Fig 21) 

(Fig 20) 

 

Note: 

Always charge your batteries in well ventilated areas. 

The charger is intended for indoor use only. Protect from moisture. 

For maximum performance , it is recommended that you replace both 

batteries at the same time if the batteries are weak. 

If the chair will not be used for a long period of time, arrange to have the 

batteries fully charge for at least once every month. 

 

According to the battery type and condition of the batteries, batteries 

usually can be fully charged in 4-10 hours. This will be indicated when the status 

light in the battery charger side panel turns green.Charging the battery longer 

than necessary will not harm the battery. We recommended that you charge the 

batteries for 8 to 10 hours after daily use. Do not charge the batteries for more 

than 24 hours. 

 

Note: There is a battery circuit diagram labeled on the frame. Please refer this 

diagram before you assemble the battery. 

 

 

 

 

Summary of Contents for P17C

Page 1: ...1 USERS MANUAL P17C HEARTWAY MEDICAL PRODUCTS CO LTD Part no 70030161...

Page 2: ...d to protect it during shipping After unpacking please check whether you have received the following main components as our standard specification See Fig 1 1 Main frame with motor 5 Armrest left 2 Se...

Page 3: ...t in excessive current draw from the batteries increased tire wear and the rapid wearing of the gear boxes and motors 7 To brake in an emergency simply release the joystick Ramps and Curbs 8 When driv...

Page 4: ...While Under The Influence Of Medication Or Alcohol 20 Check with your physician if you are taking any medication that may affect your ability to operate your power wheelchair safely 21 Do not operate...

Page 5: ...2 Be aware of nearby transmitters such as radio or TV stations and try to avoid coming close to them 3 If unintended movement or brake release occurs turn the powered wheelchair off as soon as it is s...

Page 6: ...cause electronic malfunction or promote premature corrosion of electrical components and frame B Temperature Some of the parts of the power wheelchair are susceptible to change in tempera ture The con...

Page 7: ...ke sure that tighten the tension knob in the front connection point See Fig3 Fig 2 Fig 3 Front connection point 2 Installing the Armrest Insert the armrest frame with left 5 and Right 6 each into the...

Page 8: ...s Assemble the controller 4 on right armrest 6 bracket with 2 screws See Fig 9 There is 1 plug 9 pins in the other end on the controller Please plug in the main frame 1 plugs and make sure tightly con...

Page 9: ...A Armrest Width Adjustment Loosen the tension knob slide the armrest frame to your desired width tighten the tension knob See Fig 12 Fig 12 Fig 13 B Footplate Angle Adjustment Take off the E Clip sli...

Page 10: ...adjust the backrest position to your desired angle See Fig 15 Note There are 5 different angles range when you slightly adjust the backrest to fit yours comfortable seating from the 105 to 135 E Head...

Page 11: ...your vehicle safely with the controller See Fig 17 Fig 17 SPEED CONTROL A Driving 1 Controller ON OFF Switch Press the ON OFF button I O switch located in front of the joystick to activate your power...

Page 12: ...hen the wheelchair is in operation the surface of the charger will become slightly hot In case of emergency let go of the joystick and the chair will come to a stop B Controller Display The controller...

Page 13: ...n a steady state i e Joystick is released to the neutral position even when the chair is on a slope The Electromagnetic Brakes will also be set when ever the power wheelchair is OFF but the motor leve...

Page 14: ...r extra safety When the batteries and motors are heavily strained e g from excessive loads the main circuit breaker will trip to prevent damage to the motor and the electronics If the circuit breaker...

Page 15: ...t designed to handle a long deep discharge and also are unsafe for use in power chairs The useful life of a battery is quite often a reflection of the care it receives CHARGER The battery charger take...

Page 16: ...s at the same time if the batteries are weak If the chair will not be used for a long period of time arrange to have the batteries fully charge for at least once every month According to the battery t...

Page 17: ...to your power wheelchair Do not over tighten any screws 3 Check for proper tire inflation Your power wheelchair comes with standard flat free solid tires If your power wheelchair come with optional a...

Page 18: ...ller clean while protecting it from rain or water Never hose off your power wheelchair or place it in direct contact with water 2 Keep wheels free from lint hair sand and carpet fibers 3 Visually insp...

Page 19: ...ction Check the left motor connection 3 The left motor has a short circuit to a battery connection Contact your local dealer 4 The right motor has a bad connection Check the right motor connection 5 T...

Reviews: