Noise level Test Information
Test Method(s):
The evaluator observed the sound levels from various locations around the equipment
and from a location representing the typical operator location. The equipment was
operated and cycled in its worst-case laboratory condition. Measurements were taken
20 inches away from the front and right side of the equipment at a height of 45 inches
from the floor. Measurements at the left and rear of the equipment were restricted due
to proximity with the location of the equipment relative to the walls.
Acceptable results
: Maximum sound pressure level does not exceed 80dB(A).
Test Results:
PASS
Ambient – 68.8dB(A) –
Note: Operation of the system adds no measurable sound pressure to the local
environment.
Location:
Front:
68.6dB(A)
Left Side:
68.8dB(A)
Right Side: 68.7dB(A)
Rear:
68.8dB(A)
The energy consumption for idle, average, and peak conditions varies from system to
system and depends on how the customer uses the system. For daily energy
consumption information, reference the Quality Control documentation.