Service and Maintenance
6–8
Edition September 2000
ABC
Replacing the Glass Plate
If the glass plate in your scanner is scratched or
broken, you can replace it yourself. Order No. for
the glass plate: 05355036
When reinstalling, proceed as follows:
1. Switch off the device.
Warning:
Pull out the power plug.
The ends of the fluorescent lamps can heat up to
100 °C when switched on.
2. Open the scanner lid.
3. Loosen the two slotted screws at the
front and the slotted screw at the rear
right for the glass plate in the scanner.
Remove the three mounting posts in
which the screws are located.
The
mounting post at the rear left does not
have to be loosened.
4. Remove the glass plate.
Carefully remove the glass plate by
putting your fingers into the cutouts for
the front mounting post and raising the
glass plate.
5. Carefully insert the new glass pane.
Carefully slide the pane against the stop
pins to the rear and left. Screw the
3 mounting posts back into place. When
doing so
make absolutely certain
that
the mounting posts are inserted the right
way round to ensure that the glass pane
does not break when the screws are
tightened.
3
Removed
mounting
post