Working with the Nexscan F 4100/F 4200
Nexscan F 4100/F 4200 – Operation
4–17
ABC
Performing a Basic Level Adjustment (for Newcolor)
Basic level adjustment is performed by Heidelberg
and normally does not have to be repeated.
It is possible that a base level adjustment becomes
necessary as part of an installation procedure or
that a particular error message at the monitor
requires it. In these cases, proceed as follows:
Remove the scanning tray from the Scanner. Clean
the glass surface. It must be absolutely clean,
especially the white balance strip.
Select the menu item
Scan
>
Special scanner
functions
...