Commissioning | Configuring the axes
7
MM
X
24.9979
80.0004
50.0020
79.9991
75.0001
79.9985
100.0010
80.0002
Explanation
The following overview explains the values that you can define individually. All
values not listed here must be copied as is from the example. The values must be
separated by tab stops.
Value
Explanation
Value
Explanation
MM
Unit of measurement:
millimeters (alternatively:
IN for inches)
X
Alignment axis
(X or Y)
25.0
Spacing of supporting
points on the X axis
20.0
Spacing of supporting
points on the Y axis
5
Number of supporting
points on the X axis
5
Number of supporting
points on the Y axis
0.0000
Actual value of the first
supporting point on the X
axis
0.0000
Actual value of the first
supporting point on the Y
axis
25.0012
Actual value of the second
supporting point on the X
axis
-0.0010
Actual value of the second
supporting point on the Y
axis
The file contains an additional row with X and Y values for each
supporting point.
104
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
Summary of Contents for GAGE-CHEK 2000 Series
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Page 12: ......
Page 13: ...1 Fundamentals ...
Page 22: ......
Page 23: ...2 Safety ...
Page 28: ......
Page 29: ...3 Transport and storage ...
Page 34: ......
Page 35: ...4 Mounting ...
Page 41: ...5 Installation ...
Page 51: ...6 Basic operation ...
Page 74: ......
Page 75: ...7 Commissioning ...
Page 119: ...8 Setup ...
Page 154: ......
Page 155: ...9 Measuring ...
Page 169: ...10 File management ...
Page 176: ......
Page 177: ...11 Settings ...
Page 210: ......
Page 211: ...12 Servicing and maintenance ...
Page 220: ......
Page 221: ...13 What to do if ...
Page 225: ...14 Removal and disposal ...
Page 227: ...15 Specifications ...