Commissioning | Configuring the axes
Explanation
The following overview explains the parameters and values that you can define
individually. All items not listed here must be copied as is from the example.
Group
Parameters and values (example)
Explanation
<element id="x">
25
</element>
Spacing of supporting points on the
X axis. In this example: 25 mm
<group
id="
CellSize
">
<element id="y">
20
</element>
Spacing of supporting points on the
Y axis. In this example: 20 mm
<element id="x">
5
</element>
Number of supporting points on the
X axis. In this example: 5 supporting
points
<group
id="
GridSize
">
<element id="y">
5
</element>
Number of supporting points on the
Y axis. In this example: 5 supporting
points
<element id="0-0" Y="
0
" X="
0
"/>
Actual value of the first supporting
point in mm. In this example:
X = 0
Y = 0
<element id="1-0" Y="
-0.001
"
X="
25.001200000000001
"/>
Actual value of the supporting point in
mm. In this example:
X = -0.001
Y = 25.001200000000001
<group
id="
Level0
">
For each supporting point, the group contains an additional element with these
parameters.
7
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
109
Summary of Contents for GAGE-CHEK 2000 Series
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Page 12: ......
Page 13: ...1 Fundamentals ...
Page 22: ......
Page 23: ...2 Safety ...
Page 28: ......
Page 29: ...3 Transport and storage ...
Page 34: ......
Page 35: ...4 Mounting ...
Page 41: ...5 Installation ...
Page 51: ...6 Basic operation ...
Page 74: ......
Page 75: ...7 Commissioning ...
Page 119: ...8 Setup ...
Page 154: ......
Page 155: ...9 Measuring ...
Page 169: ...10 File management ...
Page 176: ......
Page 177: ...11 Settings ...
Page 210: ......
Page 211: ...12 Servicing and maintenance ...
Page 220: ......
Page 221: ...13 What to do if ...
Page 225: ...14 Removal and disposal ...
Page 227: ...15 Specifications ...