Settings |
Axes
11
Parameters
Explanation
Inversion of reference mark
pulses
Specifies whether the reference mark pulses are evaluated in
inverted form
Settings
ON
: Reference pulses are evaluated in inverted form
OFF
: Reference pulses are not evaluated in inverted form
Default value:
OFF
Reference point displacement
Configuration of the offset between the reference mark and the
zero point
"Reference point displacement",
11.6.10
Reference point displacement
Path:
Settings
►
Axes
►
<Axis name>
►
Encoder
►
Reference marks
►
Reference point displacement
Parameters
Explanation
Reference point displacement
Activation of offset calculation between reference mark and
datum of the machine
Setting range:
ON
or
OFF
Default value:
OFF
Reference point displacement
Manual input of the offset (in mm or degrees according to the
selected encoder type) between reference mark and datum
Default value:
0.00000
Current position for reference
point shift
Apply
applies the current position as an offset (in mm or
degrees, depending on the selected encoder model) between
the reference mark and the zero point
202
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
Summary of Contents for GAGE-CHEK 2000 Series
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Page 12: ......
Page 13: ...1 Fundamentals ...
Page 22: ......
Page 23: ...2 Safety ...
Page 28: ......
Page 29: ...3 Transport and storage ...
Page 34: ......
Page 35: ...4 Mounting ...
Page 41: ...5 Installation ...
Page 51: ...6 Basic operation ...
Page 74: ......
Page 75: ...7 Commissioning ...
Page 119: ...8 Setup ...
Page 154: ......
Page 155: ...9 Measuring ...
Page 169: ...10 File management ...
Page 176: ......
Page 177: ...11 Settings ...
Page 210: ......
Page 211: ...12 Servicing and maintenance ...
Page 220: ......
Page 221: ...13 What to do if ...
Page 225: ...14 Removal and disposal ...
Page 227: ...15 Specifications ...