19
The ppO2 value is shown with a simple
blink code. The code is repeated with up-
dated values every fi ve seconds.
The display has a resolution of 0.1bar.
All values above 1.0 bar up to 1.6 bar are
green, all values below 1.0 bar are red
blink codes. 1.0 bar itself is displayed as
1x yellow. Every sensor has its seperate
display. They show pure measurement
data and no average values.
If any sensor fails, its display does not in-
terfere with the display of the remaining
sensors
< 0.25bar:
8 x Red
0.25-0.35bar:
7 x Red
0.35-0.45bar:
6 x Red
0.45-0.55bar:
5 x Red
0.55-0.65bar:
4 x Red
0.65-0.75bar:
3 x Red
0.75-0.85bar:
2 x Red
0.85-0.95bar:
1 x Red
0.95–1.05bar:
1 x Yellow
1.05-1.15bar:
1 x Green
1.15-1.25bar:
2 x Green
1.25-1.35bar:
3 x Green
1.35-1.45bar:
4 x Green
1.45-1.55bar:
5 x Green
1.55-1.65bar:
6 x Green
> 1.65bar:
Constant red