Helmer Scientific i.Series
®
and Horizon Series
™
Freezer - Undercounter
Service and Maintenance Manual
360417/A
31
Condenser Grill
Disconnect the freezer from AC power when cleaning.
NOTICE
Some equipment surfaces may be sharp. Take care when cleaning to avoid injury.
If the freezer is located in an environment where it is exposed to excessive lint or dust, the condenser grill may require cleaning
more frequently than stated in the preventive maintenance schedule.
Clean the condenser grill using a soft brush and a vacuum cleaner.
Door Gasket
Clean with a soft cloth and mild soap and water solution.
Probe Bottle
Clean and Refill Probe Bottle
1. Remove all probes from the bottle.
2. Remove the bottle from the bracket and empty any remaining solution.
3. Clean the bottle with a 1:9 ratio of bleach to water solution or a company approved cleaner/disinfectant.
4.
Refill the bottle with 4 oz. (120 mL) of product simulation solution (1:1 ratio of water to propylene glycol).
5. Cap the bottle tightly to minimize evaporation.
6. Place the bottle in the bracket.
7. Replace the probe(s), immersing at least 2” (50 mm).
i.C³
®
Touchscreen
Clean the touchscreen with a soft, dry cotton cloth.
i.Series