9 Installation and Configuration
9.1
Location Requirements
Keep all ventilation openings in the enclosure or, in the structure for building-in, clear of obstruction.
♦ Has a grounded outlet meeting the electrical requirements listed on the product specification label.
♦
Is clear of direct sunlight, high temperature sources, heating vents, and air conditioning vents.
♦
Has a minimum of 3” (76 mm) of space behind the freezer for clearance and feature access.
♦ Meets specified limits for ambient temperature and relative humidity as stated in the Product Specifications section of
this manual.
9.2
Placement and Leveling
NOTICE
• To prevent tipping, ensure the casters (if installed) are unlocked and the door is closed before moving the freezer.
• Use of leveling feet or casters is required.
• Do not sit, lean, push or place heavy objects on top surface.
• Do not lean on or push down on an open door or extended drawers.
• To avoid damaging refrigerant tubing or risking refrigerant leak, use caution when moving or operating the unit.
1. Ensure the door is secured and casters (if installed) are unlocked.
2. Move the freezer into place. Lock the casters if installed.
3. Ensure the freezer is level.
Note
Helmer recommends the use of leveling feet and wall and floor brackets (PN 400472-2) for stabilization. Contact Helmer
Technical Service for parts and instruction.
9.3
Stacked Undercounter Units
NOTICE
•
For stacked configuration, both units must have leveling feet installed.
• Back brace bars and front stabilizing brackets must be installed (Blue - PN 400821-1; Stainless Steel - PN 400821-2).
• When stacking units, place the heavier unit on the bottom.
• Do not open multiple loaded drawers at the same time.
• Do not lean on or push down on an open door or extended drawers.
Contact Helmer or your distributor for more information regarding the stacking kit and methods to secure both units to the wall
and/or floor.