Helmer Scientific i.Series
®
and Horizon Series
™
Refrigerator - Pass-Thru
Service and Maintenance Manual
360429/A
13
3 Controls
i.Series models are equipped with the i.C
3
monitoring and control system. The i.C
3
system combines temperature control and
monitoring into a single user interface.
Note
Please refer to the i.C
3
User Guide for complete information regarding the i.C
3
User Interface.
3.1
Home Screen and Screensaver
The Home Screen is the default screen and is displayed when:
♦
The Home icon is touched from any other screen.
♦
There is no interaction for two minutes on any screen other than those used to enter a password.
Home
screen
Home
screensaver
3.2
Home Screen Functions
Note
Refer to the i.C³ User Guide for options available on all i.C³ screens.
♦
View current interior cabinet temperature readings
♦
View the current system time and date
♦ Access any of the five homescreen applications (touch
i.C³ APPS
for additional applications)
♦
View information about current alarm events
♦
View whether the monitoring system is running on battery power
♦
Mute audible alarms
♦ Turn the chamber light on and off
♦
View a graph of the chamber temperature
♦
View unit ID
♦
Shortcut to Event Log
3.3
Alarm Reference
If an alarm condition is met, an alarm activates. Some alarms are visual only; others are visual and audible. Some alarms are sent
through the remote alarm interface. The table below indicates if an alarm is audible (A), visual (V), or sent through the remote
alarm interface (R).
Table 2. i.Series Alarm Reference
Alarm
Alarm Type
Alarm
Alarm Type
High Temperature
A, V, R
Low Battery
V
Low Temperature
A, V, R
No Battery
A, V, R
Compressor Temperature
A, V, R
Probe Failure
A, V, R
Door Open (Time)
A, V, R
Communication Failure
A, V, R
Power Failure
A, V, R
i.Series