3.4 Settings
>
Through the i.C³ monitoring and control system, current settings may be viewed and changed. To view settings, touch
i.C³ APPS
,
Settings
. Use a touch-drag motion to scroll up or down to select the desired setting.
Settings screens
Note
•
If the Settings screen is password protected enter appropriate password. If viewing settings for the first time, enter factory
default password of “1234”.
• Default values for general settings, alarm settings, and display settings are available in the i.C³ User Guide.
•
Changing temperature settings affects operation of the refrigerator. Do not change settings unless instructed in product
documentation or by Helmer Technical Service.
The i.C³ temperature monitor and controller is programmed at the factory. To change a setting, first enter the Settings screen, then
select the setting. The method for accessing the Settings mode for each setting varies.
Device Control Settings
Device control settings are programmed at the factory. Setpoints can be viewed and changed through the i.C³ monitoring and
control system. To view temperature setpoints, touch
i.C³ APPS
,
Settings, Device Control Settings
.
Device Control Settings screen.
Table 3. Setpoints
Setting
Initial Factory Value
Temperature Setpoint
4.0 °C
(iBR models)
5.0 °C
(iPR models)
Upper Rail
1.5 °C
Lower Rail
-1.5 °C
Delay on Start-Up
1 minute
Speed During Probe Error
50%