Helmer Scientific i.Series
®
and Horizon Series
™
Refrigerator - Pass-Thru
Service and Maintenance Manual
360429/A
16
User Configurable Alarm Settings
The following alarm settings may be changed by the operator. The setpoint for temperature alarms may be changed (where
applicable), as well as the time delay between when the alarm condition commences and when the visual and audible alarms
are initiated.
Table 4. User Configurable Alarms
Alarm
Description
Default Value
Default
Time Delay
Primary Monitor Probe High Temp
High temperature at which alarm condition occurs
5.5 °C
(iBR models
)
0 minutes
6.5 °C
(iPR models
)
Primary Monitor Probe Low Temp
Low temperature at which alarm condition occurs
1.5 °C
(iBR models
)
0 minutes
2.0 °C
(iPR models)
Compressor High Temp
High temperature at which alarm condition occurs
50 °C
0 minutes
Power Failure
Time after power failure occurs until alarm sounds
-
1 minute
Probe Failure
Time after probe failure occurs until alarm sounds
-
0 minutes
Door Open (Time)
Time door remains open until alarm sounds
-
3 minutes
Alarm Settings screen
Change an Alarm Setting:
1. Touch
i.C³ APPS
,
Settings
.
2. Enter Settings password (default password is “1234”).
3. Scroll down and touch
Alarm Settings
.
4. Touch minus (-) or plus (+) on spin box corresponding to alarm setting to be changed.
5. Touch
Home
to exit Alarm Settings screen.
i.Series