Helmer Scientific i.Series
®
and Horizon Series
™
Refrigerator - Pass-Thru
Service and Maintenance Manual
360429/A
18
3.5
Sensor Calibration
>
Sensor calibration values are programmed at the factory. Calibration values can be viewed and changed through the i.C³
monitoring and control system. To view calibration settings, touch
i.C³ APPS
,
Settings
, and select
Sensor Calibration
.
Settings screen
Sensor Calibration screens
Notes
•
If the Settings screen is password protected enter the appropriate password. If viewing settings for the first time, enter the
factory default password of “1234”.
•
Offset values are adjusted to the tenth of a degree (0.1).
• After one hour of no interaction, the Home screen or Temperature Graph screensaver (if enabled) is displayed.
•
The Compressor Probe offset is factory-preset and should not be changed unless directed by Helmer Technical Service.
View Sensor Calibration Values:
1. Touch
Settings
icon.
2. Enter Settings password.
3. Touch
Sensor Calibration
. Sensor offset values and their current temperature readings are displayed.
4. Touch
Home
to return to Home screen.
i.Series