Helmer Scientific i.Series
®
and Horizon Series
™
Refrigerator - Pass-Thru
Service and Maintenance Manual
360429/A
21
Enter the New Offset Value:
1. Touch
Home
,
i.C³ APPS
,
Settings
.
2. Enter Settings password (default password is “1234”).
3. Touch
Sensor Calibration
.
4. Touch minus (-) or plus (+) on
Control Probe Offset
spin box.
♦ Raise offset value to lower chamber temperature; lower offset value to raise chamber temperature.
5. Touch
Home
to return to home screen.
Compressor Probe
The compressor probe has been factory-calibrated. Changing the calibration settings is not typically necessary and should not be
performed unless directed by Helmer Technical Service.
i.Series