Helmer Scientific i.Series
®
and Horizon Series
™
Refrigerator - Pass-Thru
Service and Maintenance Manual
360429/A
22
Factory Default Settings
Settings listed below may be simultaneously returned to factory default values.
Note
The factory default settings may not be the same as the settings that were factory-calibrated before the refrigerator shipped.
Table 6. Restored Settings
Setting
Restored Value
Home Screen Application Icons
i.C³ APPS, Settings, Temperature Graph, Temperature Alarm Test, Information Logs
Display Brightness
3 (brightest)
Password (for Settings screen)
1234
Sounds
On
Alarm Volume
9
Alarm Tone
3
Temperature Calibration Values
Not affected (values previously entered during setup)
Unit ID
Not affected (previously selected during setup)
Date Format
MM/DD/YYYY
Day
Not affected (maintained in real-time clock)
Month
Year
Time Format
12-hour
Minute
Not affected (maintained in real-time clock)
Hour
AM/PM
Language
Not affected (language previously selected during setup)
Temperature Units
ºC
Password Protection (for Settings screen)
On
Temperature Graph Screensaver
On
Light Off Delay (on/off)
On
Light Off Delay
5 minutes
High Temperature Alarm Setpoint
5.5 ºC (iBR models)
6.5 ºC (iPR models)
High Temperature Alarm Time Delay
0 minutes
Low Temperature Alarm Setpoint
1.5 ºC (iBR models)
2.0 ºC (iPR models)
Low Temperature Alarm Time Delay
0 minutes
Power Failure Alarm Time Delay
1 minute
Probe Failure Alarm Time Delay
0 minutes
Door Open (Time) Alarm Time Delay
3 minutes
Compressor Temperature Alarm Setpoint
50.0 ºC
Compressor Temperature Alarm Time Delay
1 minute
Chamber Setpoint
4.0 ºC (iBR models)
5.0 ºC (iPR models)
Upper Rail
1.5 ºC
Lower Rail
-1.5 ºC
Delay on Start-Up
2 minutes
Speed During Probe Error
50%
Restore Settings:
1. Touch
Settings
icon.
2. Scroll down and touch
Restore Factory Settings
button. Restore Factory Settings confirmation box appears.
3. Touch
to restore factory settings or
to maintain current settings and clear message.