8 Product Specifications
8.1
Operating Standards
These units are designed to operate under the following environmental conditions:
♦
Indoor use only
♦
Altitude (maximum): 2000 m
♦
Ambient temperature range: 15 °C to 32 °C
♦
Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C, decreasing linearly to 50% at 40 °C
♦
Temperature control range: -15 °C to -30 °C
♦
Sound level is less than 70 dB(A).
♦
EMC Environment: Basic
Table 6. Electrical Specifications (Laboratory and Plasma Storage)
104
105
120 / 125
Input Voltage and
Frequency
115 V, 60 Hz
115 V, 60 Hz; 230 V, 50 Hz;
230 V, 60 Hz
115 V, 60 Hz; 230 V, 50 Hz;
208/230 V, 60 Hz
Voltage Tolerance
±10%
Circuit Breakers
N/A
6.0 A (230 V only, quantity 2)
12.0 A (quantity 2)
Current Draw
5.75 A (115 V, 60 Hz)
5.75 A (115 V, 60 Hz)
2.9 A (230 V, 50 Hz)
3.1 A (230 V, 60 Hz)
8.5 A (115 V, 60 Hz)
3.8 A (230 V, 50 Hz)
4.3 A (208/230 V, 60 Hz)
Power Source
Grounded outlet, meeting national electric code (NEC) in the U.S. and local electrical requirements in all locations.
Remote Alarm
Capacity
i.Series: 0.5 A at 125 V (AC)
1 A at 250 V (DC)
Horizon Series: 0.25 A at 30 V (RMS)
0.25 A at 60 V (DC)
i.Series: 0.5 A at 30 V (RMS)
1.0 A at 24 V (DC)
Horizon Series: 0.25 A at 30 V (RMS);
0.25 A at 60 V (DC)
i.Series: 0.5 A at 30 V (RMS);
1.0 A at 24 V (DC)
Horizon Series: 0.25 A at 30 V (RMS);
0.25 A at 60 V (DC
* Amperage values are subject to change. Refer to the product specification label on your unit for current values.
NOTICE
• The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
• If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.