3.2 Operation
Notes
• Refer to the i.C
3
User Guide for complete information regarding the i.C
3
User Interface.
• The i.C³ Home screen displays temperature and alarm information, and provides icons to gain access to other functions
of the i.C³.
• After two minutes of inactivity, the screensaver will be displayed. To return to the Home screen, touch the screensaver.
Home screen
Home screensaver (touch to return to Home screen)
3.3
Change Temperature Setpoint
Note
The Temperature Setpoint toggle button can be accessed from either the initial Settings screen or the Device Control
Settings screen.
>
> Enter the Settings password. Select Temperature setpoints. Touch minus (-) or plus (+) on the spin box to
change the value.
Settings
screen
Device
Control
Settings
screen
Notes
• Default Settings password is 1234.
• Default setpoint is -30.0 °C.
Helmer Scientific i.Series
™
-