3.5
Active Alarms
Home screen with active alarm
Table 1. i.Series Active Alarms
Alarm
Description
Primary Monitor Probe High Temp
Primary monitor probe reading is above high temperature alarm setpoint
Primary Monitor Probe Low Temp
Primary monitor probe temperature reading is below low temperature alarm setpoint
Primary Monitor Probe Failure
Primary monitor probe not functioning properly
Secondary Monitor Probe High Temp
(if installed)
Secondary monitor probe reading is above high temperature alarm setpoint
Secondary Monitor Probe Low Temp
(if installed)
Secondary monitor probe temperature reading is below low temperature alarm setpoint
Secondary Monitor Probe Failure
(if installed)
Secondary monitor probe not functioning properly
Control Probe Failure
Control probe not functioning properly
Evap Defrost Probe Failure
Evaporator defrost probe not functioning properly
Compressor High Temperature
Compressor temperature reading is above high temperature alarm setpoint
Compressor Probe Failure
Compressor probe not functioning properly
Power Failure
Power to unit has been disrupted
Door Open
Door is open beyond user-specified duration
Low Battery
Rechargeable battery voltage is low
No Battery
Battery is not connected
Communication Failure Messages
1, 2, 3
1 Communication lost between i.C³ display board and control board
2 Communication lost between i.C³ display board and internal system memory
3 Corrupt database
Helmer Scientific i.Series
™
-
Horizon Series
™
Freezer
Instructions for Use
360380-ENG/B
10