background image

6.4 

Set Parameter Values

1.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to enter program mode. 

2.  The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3.  Press and release 

SEL

 button to scroll through the parameters.

4.  Once the desired parameter is selected, press and hold the 

SET

 button while pressing the 

Up

 or 

Down Arrow

 to select 

the desired value.

5.  Release 

SET

 button.  The new setting is saved.

6.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to exit program mode.

  Note

Contact Helmer Technical Service to set Hysteresis values.

Table 4.  Parameter Values

Parameter

Visual Indicator

Range

Default

Celsius or Fahrenheit

None

.C, .F

.C

High Temperature

MONITOR Lamp & HIGH Lamp

-40.0 to 40.0 (°C); -40 to 104 (°F)

-20.0°C

Low Temperature

MONITOR Lamp & LOW Lamp

-40.0 to 40.0 (°C); -40 to 104 (°F)

-40.0°C

Monitor Offset

MONITOR Lamp

-10.0 to 10.0 (°C); -18 to 18 (°F)

Varies

Control Offset

CONTROL Lamp

-10.0 to 10.0 (°C);-18 to 18 (°F)

Varies

Hysteresis

CONTROL Lamp

0.5 to 2.5 (°C); 1 to 5 (°F)

2.0°C

6.5 

Set Temperature Units

  Note

If temperature units are changed, the temperature setpoints, offsets and alarm settings must be recalibrated.

1.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to enter program mode.

2. 

The LED Display will show ˚C or ˚F to indicate Celsius or Fahrenheit.

3.  Press and hold the 

SET

 button while pressing the 

Up

 or 

Down Arrow

 to select the desired temperature unit.

4.  Release 

SET

 button. The new setting is saved.

5.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to exit program mode.

6.6  Temperature Calibration Offsets

Temperature calibration offsets indicate an acceptable margin of error between the actual temperature value and the desired 

temperature value.

Monitor Offset

 

Value is factory-set to match a calibrated reference thermometer.

 

♦ Refer to the service manual for instructions regarding changing the Monitor Offset.

Control Sensor Offset and Hysteresis

The control sensor affects the reading of the control probe temperature and therefore the actual temperature of the freezer.  

This should not be adjusted from the original setting unless directed by Helmer Technical Service.

Hysteresis helps control the refrigeration based on the control probe temperature reading and the set point and should not 
be changed from the default setting. 

NOTICE

Control Sensor Offset and Hysteresis are factory-preset and should not be changed. Contact Helmer Technical Service for 

instructions regarding changing these values.

Helmer Scientific i.Series

 -

 

Horizon Series

 

 

Freezer

Instructions for Use

360380-ENG/B

16

Summary of Contents for HLF120

Page 1: ...zon Series Upright Undercounter Laboratory i Series iLF104 ADA iLF105 iLF120 iLF125 Horizon Series HLF104 ADA HLF105 HLF120 HLF125 Plasma Storage i Series iPF104 ADA iPF105 iPF120 iPF125 Horizon Serie...

Page 2: ...emble or decompile or create or attempt to create by reverse engineering or otherwise the information from Helmer Scientific products is expressly prohibited Copyright and Trademark Copyright 2020 Hel...

Page 3: ...e Monitoring 12 5 i Series Access Control Optional 13 5 1 Set Up 13 5 2 Open Freezer with Access Control 13 6 Horizon Series Operation 14 6 1 Initial Power Up 14 6 2 Display Minimum and Maximum Monito...

Page 4: ...roducts 1 4 Safety Symbols and Precautions Symbols found in this document The following symbols are used in this manual to emphasize certain details for the user Task Indicates procedures which need t...

Page 5: ...aminate parts prior to sending for service or repair Contact Helmer or your distributor for decontamination instructions and a Return Authorization Number 1 6 General Recommendations General Use Allow...

Page 6: ...asters if installed 2 Ensure freezer is level Note Helmer recommends the use of leveling feet Contact Helmer Technical Service for parts and instruction 2 3 Stacked Undercounter Units NOTICE For stack...

Page 7: ...nt or the chamber temperature to display higher or lower than the actual temperature Primary Monitor Probe The primary monitor probe bottle is located at the top left side of the freezer Upright probe...

Page 8: ...emperature chart recorder is available for up to 14 hours Note If chart recorder is operated on battery power the battery should be replaced to ensure the back up source has proper charge Prior to Use...

Page 9: ...tylus begins to move right release button 7 Confirm stylus is marking on paper and stops at the correct temperature 8 Calibrate chart recorder to match primary temperature if needed and close recorder...

Page 10: ...The i C3 will take approximately 2 5 minutes to boot up Start screen The language screen is displayed when the i C3 is powered on Use the Language screen to select the i C3 display language Language s...

Page 11: ...aver Home screen Home screensaver touch to return to Home screen 3 3 Change Temperature Setpoint Note The Temperature Setpoint toggle button can be accessed from either the initial Settings screen or...

Page 12: ...ouch minus or plus on spin box to set each alarm parameter Settings screen Alarms screens Alarm settings control the circumstances and timing of alarm condition indicators displayed on the i C Home sc...

Page 13: ...itor Probe Failure if installed Secondary monitor probe not functioning properly Control Probe Failure Control probe not functioning properly Evap Defrost Probe Failure Evaporator defrost probe not fu...

Page 14: ...ature Graph Upload Save Event Log Alarm Test Access Control Cancel Mute Information Logs Access Log Back Arrow Reset Contact Information Contact Helmer Defrost Cycle Scroll Zoom Information Display Br...

Page 15: ...Touch the Reset icon to the right of the display to manually reset Notes The Min Max temperature display can be turned On or Off through Display Settings Once the time reaches the maximum display of...

Page 16: ...re and manage user specific accounts to allow controlled access to the freezer Access Set Up Access Control Setup Password screen Access Control Setup screen Enter the supervisor PIN to set up Access...

Page 17: ...high alarm temperature setpoint is being changed C LOW TEMP lamp Indicates when the freezer is in a low temperature alarm condition Also indicates the low alarm temperature setpoint is being changed...

Page 18: ...record is 99 59 99 hours and 59 minutes 1 Press and hold either the Up or Down Arrow button for 1 second or 2 While the display is flashing the HI or LO value press and hold the SET button for 1 seco...

Page 19: ...simultaneously for 3 seconds to enter program mode 2 The LED Display will show C or F to indicate Celsius or Fahrenheit 3 Press and hold the SET button while pressing the Up or Down Arrow to select th...

Page 20: ...ading is above high temperature alarm setpoint Low Temperature LOW TEMP lamp flashes Chamber temperature reading is below low temperature alarm setpoint Power Failure PoFF appears on display Power to...

Page 21: ...Note The master code cannot be deleted and should be changed to prevent unauthorized user code setup Enter unique user codes for up to 100 users Each user code is stored with a specific record locatio...

Page 22: ...V 60 Hz Power Source Grounded outlet meeting national electric code NEC in the U S and local electrical requirements in all locations Remote Alarm Capacity i Series 0 5 A at 125 V AC 1 A at 250 V DC...

Page 23: ...481 219 230V 50 Hz 3 8 208 230 V 60 Hz 4 3 HLF125 115V 60 Hz 8 5 25 708 Upright Single hinged solid 4 30 75 x 80 x 38 5 782 x 2032 x 978 478 217 230V 50 Hz 3 8 208 230 V 60 Hz 4 3 Table 8 Laboratory...

Page 24: ...lid 8 30 75 x 80 x 38 5 782 x 2032 x 978 554 252 230V 50 Hz 3 8 208 230 V 60 Hz 4 3 Table 10 Plasma Storage Freezer Specifications Undercounter Model Voltage Code Amps Cu Ft Liters Cabinet Door Drawer...

Page 25: ...2 19 EU Waste Electrical and Electronic Equipment WEEE When disposing of this product in countries affected by this directive Do not dispose of this product as unsorted municipal waste Collect this pr...

Page 26: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2020 Helmer Inc 360380 ENG B...

Reviews: