2 Installation
2.1 Location
♦ Has a grounded outlet meeting the electrical requirements listed on the product specification label.
♦
Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents.
♦
Upright units require minimum 8” (203 mm) above, and minimum 3” (76 mm) behind.
♦
Undercounter units require minimum 3” behind the unit for clearance and feature access.
♦ Meets limits specified for ambient temperature and relative humidity.
2.2
Placement and Leveling
NOTICE
• To prevent tipping, ensure the casters (if installed) are unlocked and the door is closed before moving the freezer.
• Do not use the water evaporation tray, located on the rear of the upright freezer, as a handle. The tray may be hot.
• Do not sit, lean, push or place heavy objects on top surface of undercounter units.
1. Roll freezer into place. Lock casters if installed.
2. Ensure freezer is level.
Note
Helmer recommends the use of leveling feet. Contact Helmer Technical Service for parts and instruction.
2.3
Stacked Undercounter Units
NOTICE
•
For stacked configuration, both units must have leveling feet installed.
• Back brace bars and front stabilizing brackets must be installed (Blue - PN 400821-1; Stainless Steel - PN 400821-2)
• When stacking units, place the heavier unit on the bottom.
• Do not open multiple loaded drawers or baskets at the same time.
Contact Helmer or your distributor for more information regarding the stacking kit and methods to secure both units to the wall
and / or floor.
2.4
AC Power Cord
Use manufacturer supplied power cord only.
Install power cord
If packaged with modular cord, insert plug securely into the freezer power receptacle prior to connecting to grounded outlet.
Ensure power cord has been secured with strain relief zip tie.