Helmer Scientific i.Series
®
and Horizon Series
™
Freezer - Undercounter
Operation Manual
360376/C
14
5 i.Series
®
Access Control (Optional)
Allows user-specific secure access to the freezer.
Notes
• During a power failure, the optional Access Control lock will remain locked until battery power is depleted or until the
back-up battery key switch is switched OFF.
•
During a power failure, switching the back-up battery key switch OFF will disable the monitoring system.
•
During a power failure, switch the battery back-up switch OFF and use the mechanical door key to provide secure
storage for freezer contents.
• Refer to the i.C
3
User Guide for complete information regarding Access Control.
5.1 Setup
Configure and manage user-specific accounts to allow controlled access to the freezer.
>
> Access Setup
Access Control Setup Password screen
Access Control Setup screen
Enter the supervisor PIN to set up Access Control and follow the on-screen prompts to set up users.
Notes
• Initial factory supervisor PIN = 5625
• The supervisor PIN can not be deleted, and should be changed to prevent unauthorized user ID setup. The supervisor
PIN does not allow access to the unit. At least one user ID must be set up to gain access to the unit.
5.2
Open Freezer with Access Control
Access Control keypad
Enter a valid PIN using the keypad.