6.7
Active Alarms
The controller displays temperature and alarm information.
Table 4. Horizon Series Active Alarms
Alarm
Visual Indicator
Description
High Temperature
HIGH TEMP lamp flashes
Chamber temperature reading is above high
temperature alarm setpoint
Low Temperature
LOW TEMP lamp flashes
Chamber temperature reading is below low
temperature alarm setpoint
Power Failure
”PoFF” appears on display
Power to unit has been disrupted
Probe Failure
“Prob” appears on display
Probe not functioning properly
Door Open < 3 min.
DOOR ALARM lamp lights
Door is open (less than three minutes)
Door Open > 3 min.
DOOR ALARM lamp flashes
Door has been open 3 minutes or longer*
*Audible alarm will sound after door is open for 3 minutes.
6.8
Mute and Disable Audible Alarms
Note
Muting audible alarms does not disable alarm lamps or signals sent through the remote alarm interface.
♦
Press
Down Arrow
(Mute) to mute audible alarms.
♦
To disable all audible alarms, insert the key in the Alarm Disable switch and turn.