360142-D/D
10
General Information
3.4.2
Install Chart Paper
1
Press and hold
C
button. When stylus begins to move left, release button. The LED flashes to
indicate current temperature range.
2
When stylus stops moving, remove chart knob then move knob up and away.
3
Place chart paper on chart recorder.
4
Gently lift stylus and rotate paper so current time line corresponds to time line groove.
5
Hold chart paper and reinstall chart knob.
NOTE
For accurate temperature reading, ensure that current time is aligned with time line
groove when chart knob is tightened.
6
Confirm temperature range is set to the correct value.
7
Press and hold
C
button. When stylus begins to move right, release button.
8
Confirm stylus is marking temperature correctly.